Accuracy and reproducibility


Cell research is critical across multiple fields, where precise calculation of cell count and viability is a key of step. According to GB/T42076.1-2022, both the hemocytometer method and image analysis method are current national standard methods for cell counting. The efficiency of cell counting (usually measured by the time taken per sample), accuracy (typically benchmarked against manual counting), and reproducibility (commonly evaluated by the coefficient of variation, CV, from multiple measurements of the same sample) are essential parameters for cell counting devices. Newtonoptic cell counter utilizes automated image acquisition combined with proprietary algorithms to deliver test results such as cell concentration and viability within one second, providing highly reproducible data. This enables users to rapidly obtain accurate experimental data.

 

Accuracy

Target

Multiple cell samples with known concentrations were obtained by performing serial dilutions of a cell sample with a known concentration. These samples were measured using the cell counter to validate the accuracy of the cell counting results and the linearity of dilution within this concentration range.

Experimental protocol

1After harvesting and centrifuging the HEK293 cells, the cell pellet was resuspended in PBS. Manual counting determined the concentration of the cell suspension to be 8.606 × 10 cells/mL. The sample was labeled as Cell1

2The Cell1 suspension was serially diluted with PBS at 2-fold, 4-fold, 8-fold, and 16-fold dilutions to obtain Cell2, Cell3, Cell4, and Cell5, respectively.

3Samples Cell1 through Cell5 were measured using the Newtonoptic Tunin cell counter to determine cell concentration and viability.

Result

Correlation/Fitting analysis between the measured and theoretical cell concentrations

 

Conclusion

A comparison between the theoretical and measured concentrations demonstrated that, within the cell concentration range of 5 × 10 to 8 × 10 cells/mL, the Newtonoptic cell counter provided accurate results with good dilution linearity. The fitting of the counting results yielded an R² value greater than 0.99.

 

 

Reproducibility

Target

The reproducibility of the cell counter was validated by performing multiple measurements on HEK293 cell samples at different concentrations.

 

Experimental scheme

1Three representative concentrations of HEK293 cells  high, medium, and low  were prepared, and the samples were labeled as Cell1, Cell2, and Cell3, respectively

2The three cell concentrations were analyzed using the Newtonoptic Tunin cell counter. Each sample was measured six times to evaluate the reproducibility of the counting results

3Manual counting was performed using a hemocytometer and compared with the instrument analysis results. The results are presented below

Comparison between instrument-measured concentration and manual counting concentration

 

文本框: Medium concentration

CV values of concentration measurements for high, medium, and low cell samples

 

Conclusion

The counting results from the Newtonoptic cell counter were close to those of manual counting. The Newtonoptic cell counter demonstrated excellent reproducibility, with CV values all within 5%.